• Nexus 3001XLM

    Brinell hardness tester with fully automatic indent video measuring system, equipped with a automatic motorized turret/revolver (indentor/objective positions). Optical system with high quality objective. Powerful PC based fully automatic indent measuring software IMPRESSIONS™, with indent ZOOM function, automatic focus, automatic measurements, database testing results storage and image storage. Standard equipped with motorized spindle for fully automatic procedure.

    Some characteristics:

    - 2-Position motorized turret (indentor/objective)
    - Auto-load, -indentation, -focus, -measuring, unload
    - High resolution CCD camera
    - IMPRESSIONS™ hardness testing software for automatic measuring and data/image storage
    - Powerful PC with 15” industrial LCD touch screen, height adjustable bracket, keyboard and mouse (optional) - Test loads 30kgf/294.2N - 3000kgf/29kN

    Download PDF

  • Details

    Procedure:

    Auto-load, automatic indentation, automatic focus, automatic indent measuring, unload.

    Manual mode use:

    • LCD display showing Brinell and Vickers value, statistics and tester settings
    • Simultaneous conversion to Rockwell, Vickers, Brinell and Leeb rebound testing
    • Microscope with analogue scale for indentation measurement

  • Technical Data

    Features:

    Hardness scales

    Brinell (HB)

    Load range

    30kgf/294.2N to 3000kgf/29kN

    Test loads

    30kgf/294.2N, 31.25kgf/306N, 62.5kgf/612N, 100kgf/980.7N, 125kgf/1225N, 187.5kgf/1838N, 250kgf/2.45kN, 500kgf/4.9kN, 750kgf/7.35kN, 1000kgf/9806N, 1500kgf/14.7kN, 3000kgf/29kN

    Brinell test range HB

    30kgf/294.2N, 31.25kgf/306N, 62.5kgf/612N, 100kgf/980.7N, 125kgf/1225N, 187.5kgf/1838N, 250kgf/2.45kN, 500kgf/4.9kN, 750kgf/7.35kN, 1000kgf/9806N, 1500kgf/14.7kN, 3000kgf/29kN

    Indentors

    Balls 1mm (optional) - 2.5mm - 5mm - 10mm

    Test force selection

    Electronic, closed loop, load cell, force feedback system, indication in kgf or N. Test force selectable over menu operation

    Test procedure

    Automatic, loading/dwell/unloading

    Loading speed

    Variable, depending on test load

    Test force accuracy

    Tolerance within 0.5% full range

    User display

    Diameter of indent, length of diagonals, hardness value, converted value, test force, online statistics

    Display resolution

    0.1 HB

    Hardness conversion

    Rockwell, Vickers, Brinell, Leeb and Tensile 3 scales simultaneously

    Standards

    Meets or exceeds, ISO, ASTM, JIS (Nadcap) standards

    Statistics

    Total tests, max, min, average, range, standard deviation, all in real time after each test

    Control panel

    Start test, stop test, dwell time, print, clear, menu operation for date, time, scale and load settings, language

    Firmware

    V2.01 German, English, French (standard) V2.02 English, Italian, Spanish

    Memory

    Large memory for testing results

    Connectivity

    RS-232, Bi-Directional, USB

    Loading mechanism

    Fully automatic, closed loop, force feedback, loading, dwell, unloading

    Dwell time setting

    Default 10 seconds, user defined 1 to 99 seconds

    Printer

    Optional silent high speed printer

    Optical measurement system

    Video measuring, 0.1µm reading

    Workpiece accommodation
    Vertical capacity

    335mm

    Horizontal capacity

    220mm (from center-line),

    Operating temperature

    10°C to 35°C, non-condensing

    Humidity

    10% to 90% non-condensing

    Machine dimensions

    190mm x 617mm x 985mm (XL/XLM model)

    Machine weight

    160kg

    Power supply

    100V AC to 240V AC, 50Hz/60Hz, single phase

  • Title