Nexus 3001XLM
Brinell hardness tester with fully automatic indent video measuring system, equipped with a automatic motorized turret/revolver (indentor/objective positions). Optical system with high quality objective. Powerful PC based fully automatic indent measuring software IMPRESSIONS™, with indent ZOOM function, automatic focus, automatic measurements, database testing results storage and image storage. Standard equipped with motorized spindle for fully automatic procedure.
Some characteristics:
- 2-Position motorized turret (indentor/objective)
- Auto-load, -indentation, -focus, -measuring, unload
- High resolution CCD camera
- IMPRESSIONS™ hardness testing software for automatic measuring and data/image storage
- Powerful PC with 15” industrial LCD touch screen, height adjustable bracket, keyboard and mouse (optional) - Test loads 30kgf/294.2N - 3000kgf/29kNDownload PDF
Details
Procedure:
Auto-load, automatic indentation, automatic focus, automatic indent measuring, unload.Manual mode use:
• LCD display showing Brinell and Vickers value, statistics and tester settings
• Simultaneous conversion to Rockwell, Vickers, Brinell and Leeb rebound testing
• Microscope with analogue scale for indentation measurement
Technical Data
Features:
Hardness scales
Brinell (HB)
Load range
30kgf/294.2N to 3000kgf/29kN
Test loads
30kgf/294.2N, 31.25kgf/306N, 62.5kgf/612N, 100kgf/980.7N, 125kgf/1225N, 187.5kgf/1838N, 250kgf/2.45kN, 500kgf/4.9kN, 750kgf/7.35kN, 1000kgf/9806N, 1500kgf/14.7kN, 3000kgf/29kN
Brinell test range HB
30kgf/294.2N, 31.25kgf/306N, 62.5kgf/612N, 100kgf/980.7N, 125kgf/1225N, 187.5kgf/1838N, 250kgf/2.45kN, 500kgf/4.9kN, 750kgf/7.35kN, 1000kgf/9806N, 1500kgf/14.7kN, 3000kgf/29kN
Indentors
Balls 1mm (optional) - 2.5mm - 5mm - 10mm
Test force selection
Electronic, closed loop, load cell, force feedback system, indication in kgf or N. Test force selectable over menu operation
Test procedure
Automatic, loading/dwell/unloading
Loading speed
Variable, depending on test load
Test force accuracy
Tolerance within 0.5% full range
User display
Diameter of indent, length of diagonals, hardness value, converted value, test force, online statistics
Display resolution
0.1 HB
Hardness conversion
Rockwell, Vickers, Brinell, Leeb and Tensile 3 scales simultaneously
Standards
Meets or exceeds, ISO, ASTM, JIS (Nadcap) standards
Statistics
Total tests, max, min, average, range, standard deviation, all in real time after each test
Control panel
Start test, stop test, dwell time, print, clear, menu operation for date, time, scale and load settings, language
Firmware
V2.01 German, English, French (standard) V2.02 English, Italian, Spanish
Memory
Large memory for testing results
Connectivity
RS-232, Bi-Directional, USB
Loading mechanism
Fully automatic, closed loop, force feedback, loading, dwell, unloading
Dwell time setting
Default 10 seconds, user defined 1 to 99 seconds
Printer
Optional silent high speed printer
Optical measurement system
Video measuring, 0.1µm reading
Workpiece accommodation
Vertical capacity
335mm
Horizontal capacity
220mm (from center-line),
Operating temperature
10°C to 35°C, non-condensing
Humidity
10% to 90% non-condensing
Machine dimensions
190mm x 617mm x 985mm (XL/XLM model)
Machine weight
160kg
Power supply
100V AC to 240V AC, 50Hz/60Hz, single phase
Title