• Nexus 3001

    Top quality Brinell hardness tester with various options for optical measuring systems. Ability to measure indentations starting from 30kgf/294.2N - 3000kgf/29kN. Advanced user interface with “key in” option for indent dimensions, automatic Brinell hardness calculation. Display automatically shows Brinell hardness with conversion to Rockwell, Vickers and Leeb.

    Some characteristics:

    - Load cell, closed loop system
    - Microscope with analogue scale for indentation measurement (3001 models)
    - Brinell microscope with LED ring light (3001 models)
    - Test loads 30kgf/294.2N - 3000kgf/29kN

    Download PDF

  • Details

    Standard delivery

    • Installation and user manual
    • INNOVATEST® certificate
    • Four adjustable feet
    • Hardness test block, 150-250 HBW 10/3000
    • Hardness test block, 75-125 HBW 10/1000
    • Hardness test block, 150-250 HBW 2.5/187.5
    • Analogue measuring microscope with 20x (3001/XL/XLM series)

  • Technical Data

    Features:

    Hardness scales

    Brinell (HB)

    Load range

    30kgf/294.2N to 3000kgf/29kN

    Test loads

    30kgf/294.2N, 31.25kgf/306N, 62.5kgf/612N, 100kgf/980.7N, 125kgf/1225N, 187.5kgf/1838N, 250kgf/2.45kN, 500kgf/4.9kN, 750kgf/7.35kN, 1000kgf/9806N, 1500kgf/14.7kN, 3000kgf/29kN

    Brinell test range HB

    30kgf/294.2N, 31.25kgf/306N, 62.5kgf/612N, 100kgf/980.7N, 125kgf/1225N, 187.5kgf/1838N, 250kgf/2.45kN, 500kgf/4.9kN, 750kgf/7.35kN, 1000kgf/9806N, 1500kgf/14.7kN, 3000kgf/29kN

    Indentors

    Balls 1mm (optional) - 2.5mm - 5mm - 10mm

    Test force selection

    Electronic, closed loop, load cell, force feedback system, indication in kgf or N. Test force selectable over menu operation

    Test procedure

    Automatic, loading/dwell/unloading

    Loading speed

    Variable, depending on test load

    Test force accuracy

    Tolerance within 0.5% full range

    User display

    Diameter of indent, length of diagonals, hardness value, converted value, test force, online statistics

    Display resolution

    0.1 HB

    Hardness conversion

    Rockwell, Vickers, Brinell, Leeb and Tensile 2 scales simultaneously

    Standards

    Meets or exceeds, ISO, ASTM, JIS (Nadcap) standards

    Statistics

    Total tests, max, min, average, range, standard deviation, all in real time after each test

    Control panel

    Start test, stop test, dwell time, print, clear, menu operation for date, time, scale and load settings, language

    Firmware

    V2.01 German, English, French (standard) V2.02 English, Italian, Spanish

    Memory

    Large memory for testing results

    Connectivity

    RS-232, Bi-Directional, USB

    Loading mechanism

    Fully automatic, closed loop, force feedback, loading, dwell, unloading

    Dwell time setting

    Default 10 seconds, user defined 1 to 99 seconds

    Printer

    Optional silent high speed printer

    Eyepiece microscope

    Analogue

    Workpiece accommodation
    Vertical capacity

    190mm normal model, 390mm XL/XLM model,

    Horizontal capacity

    220mm (from center-line), 220mm XL/XLM model (from center-line)

    Operating temperature

    10°C to 35°C, non-condensing

    Humidity

    10% to 90% non-condensing

    Machine dimensions

    190mm x 617mm x 790mm, 190mm x 617mm x 985mm (XL/XLM model)

    Machine weight

    130kg, 160kg XL/XLM model

    Power supply

    100V AC to 240V AC, 50Hz/60Hz, single phase

  • Title